Stationary anode X-ray tubes are used
for dental systems and small general systems such
as mobile, mobile C-arm and portable systems.
Stationary anode X-ray tubes have two-electrode tubes,
three-electrode (self bias focusing) tube, single-focus
and double-focus tubes. Also tubes with insulation
cylinder or insulation and lead cylinder are available,
so X-ray tube head can get light and compact. |

Another window opens |
Type
(1) |
Focal Spot
(mm)
|
Input
Energy
at 1sec
(W) |
Max.
Voltage
(kV) |
Max.
Current
(mA) |
Circuit
(2) |
Target Angle
(degree) |
Anode |
Dimension
Lengthx
Diameter
(3)
(mm) |
Hu
(kJ) |
Cooling Rate
(W) |
| D-041 |
0.4 |
430 |
70 |
9 |
C |
12.5 |
4.3 |
100 |
74x31
S:G74x38
SB:G74x42 |
| D-051 |
0.5 |
1750 |
100 |
22 |
C |
5 |
28 |
265 |
150x52
S:G150x60 |
| D-054 |
0.5 |
1750 |
100 |
22 |
C |
5 |
35 |
250 |
146x45
S:G146x54
SB:G146x58 |
| D-063R |
0.6 |
2020 |
110 |
20 |
C |
8 |
34 |
600 |
151x52
S:G151x60 |
| D-0711 |
0.7 |
940 |
70 |
19.9 |
C |
16 |
7 |
210 |
74x31
S:G74x38
SB:G74x42 |
| D-081B |
0.8 |
600 |
65 |
19 |
S |
20 |
6 |
128 |
69x31
S:G77x38 |
| D-082B |
0.8 |
600 |
70 |
19 |
S |
20 |
7 |
210 |
102x31
S:G107x38 |
| D-125 |
1.2 |
2700 |
125 |
40 |
C |
16 |
35 |
250 |
146x45 |
| D-183B |
1.8 |
1420 |
115 |
22 |
S |
20 |
14 |
180 |
111x41
S:G130x48 |
| D-205B |
2 |
3300 |
100 |
70 |
C |
20 |
28 |
265 |
130x52
S:G130x60 |
| D-205C |
2 |
2000/3000 |
125 |
40/80 |
S/SW |
20 |
35 |
315 |
140x52 |
DG-073B
-AC |
0.7 |
397 |
70 |
8 |
SB |
20 |
7 |
210 |
67x31 |
| DF-151R |
1.5/0.5 |
3200/680 |
110 |
60/15 |
C |
16 |
35.5 |
600 |
166x51
S:G166x60
SB:G166x64 |
Notes
1. DG : FSelf-bias tube , DF : Double Focus tube
2. All tubes are center-grounded
S
-> Self rectified , SW -> Single phase full
wave rectified , C-> Constant potential ,
CB
-> Constant potential with self-bias , SB ->
Self rectified with self-bias
3. None -> Insert tube only , S -> with Insulator Sylinder ,
SB -> with Insulator and Lead Sylinder |
|