Ionization Chambers
ionization Chambers Ionization Chamber is a detector which uses ionization caused by interaction between X-ray or gamma-ray and internal gas.
This detector is suitable for measuring in high radiation environment.

Features
1) Long life and high stability
2) Shock-resistant structure

Specification
TUBE TYPE E6854 E6861 E6866A E6866C M4952F
General Specifications  
   Fill Gas 100%Xe
   Housing Material SGP Al SUS304 SGP
   Gas pressure approx. 0.8MPa approx. 0.8MPa approx. 1.5MPa approx. 3.0MPa approx. 0.4MPa
   Length 209mm 220mm 174.5mm 220mm
   Diameter 113mm 50mm 15mm 140mm
   Effective Length 150mm 167mm 100mm 50mm
   Operating Temperature -20°C~70°C
Window Specifications  
   Material Beryllium - - SUS304
   Thickness 2.0mm 1.0mm 0.5mm 0.5mm
   Dimention Ø70mm - - Ø132mm
Electrical Specifications  
   Operating Voltage range 200~700VDC 300~700VDC 300~700VDC 100~300VDC
   Maximum Voltage 1,500VDC