Computed Tomography
![]()
An LM-CT tube is a compact ultra-high cooling X-ray tube that uses hydrodynamic bearings in its rotation mechanism and an all-metal (tungsten and molybdenum) high capacity anode.
Hydrodynamic bearings contribute lower vibration, lower noise, and greater reliability in terms of all-metal anode rotation. Temperature distribution uniformity in all-metal anodes enhances radiation cooling.
Lineup
| Type | Focal Spot (mm) |
Max. Rating (kW) |
Max. Voltage (kV) |
Max. Current (mA) |
Circuit (1) |
Angle (degree) |
Anode | |||
|---|---|---|---|---|---|---|---|---|---|---|
| Dia (mm) |
HU (kJ) |
Cooling Rate (W) |
Rotation (min-1) |
|||||||
| E7804X | 0.9*1.3 1.7*1.6 |
30/48 | 135 | 300/400 | C | 7 | 140 | 2840 | 5700 | - |
| E7830X | 0.9*0.7 1.4*1.4 |
30/48 | 135 | 300/400 | C | 7 | 140 | 2840 | 5700 | - |
♦Notes
- All tubes are cathode-grounded; C--Constant voltage