TETD Top Page > Product Lineup > X-ray Tubes > Computed Tomography

Computed Tomography

An LM-CT tube is a compact ultra-high cooling X-ray tube that uses hydrodynamic bearings in its rotation mechanism and an all-metal (tungsten and molybdenum) high capacity anode.

Hydrodynamic bearings contribute lower vibration, lower noise, and greater reliability in terms of all-metal anode rotation. Temperature distribution uniformity in all-metal anodes enhances radiation cooling.

Lineup

Type Focal
Spot
(mm)
Max.
Rating
(kW)
Max.
Voltage
(kV)
Max.
Current
(mA)
Circuit
(1)
Angle
(degree)
Anode
Dia
(mm)
HU
(kJ)
Cooling
Rate
(W)
Rotation
(min-1)
E7804X 0.9*1.3
1.7*1.6
30/48 135 300/400 C 7 140 2840 5700 -
E7830X 0.9*0.7
1.4*1.4
30/48 135 300/400 C 7 140 2840 5700 -

♦Notes

  1. All tubes are cathode-grounded; C--Constant voltage