TETD Top Page > Search by Application > General

Dental/Medical [General]

X-ray Tubes

Dental/Medical
Type
(1)
Focal Spot
(mm)
Input
Energy
at 1sec
(W)
Max.
Voltage
(kV)
Max.
Current
(mA)
Circuit
(2)
Target Angle
(degree)
Anode Dimension
Length x
Diameter
(3)
(mm)
Hu
(kJ)
Cooling Rate
(W)
D-063R 0.6 2020 110 20 C 8 34 600 151x52
S;151x60
D-125 1.2 2700 125 40 C 16 35 250 146x45
D-183B 1.8 1420 115 22 S 20 14 180 111x41
S;130x48
D-205B 2 3300 100 70 C 20 28 265 130x52
S;130x60
D-205C 2 2000/3000 125 40/80 S/SW 20 35 315 140x52
DF-151R 1.5/0.5 3200/680 110 60/15 C 16 35.5 600 166x51
S;166x60
SB;166x64

♦Notes

  1. DG: Self-bias tube; DF: Double focus tube
  2. All tubes are self-grounding
    S -- Self-rectification; SW -- Single-phase full-wave rectification; C -- Constant voltage;
    CB -- Constant voltage with self-bias; SB -- Self-rectification with self-bias
  3. Nothing -- Insert tube only; S -- with an insulating cylinder; SB -- with an insulating cylinder and a lead cylinder
General
Type Focal
Spot
(mm)
Max.
Rating
(kW)
Max.
Voltage
(kV)
Max.
Current
(mA)
Circuit
(1)
Angle
(degree)
Anode
Dia
(mm)
HU
(kJ)
Cooling
Rate
(W)
Rotation
(min-1)
E7846(2) 0.6/
1.3
11/32 130 220/
500
T,C 13.5 58 80 250 2700
E7843X 0.6/
1.2
22/50 150 370/
760
T,C 12 74 111 475 2700/
3200
E7242X 0.6/
1.5
18/50 125 290/
800
T,C 14 74 142 475 2700/
3200
E7876X 0.6/
1.5
22/54 150 300/
700
T,C 12 74 163 750 2700/
3200
E7884X 0.6/
1.2
22/54 150 300/
700
T,C 12 74 210 870 2700/
3200
E7886X 0.7/
1.3
17/400 150 280/
550
T,C 16 74 210 870 2700/
3200
E7252X 0.6/
1.2
27/75 150 400/
1000
T,C 12 74 210 475 9700
E7813X 1.0/
2.0
35/80 150 600/
1000
T,C 16 74 210 475 9700
E7255X 0.6/
1.2
40/102 150 500/
1000
T,C 12 100 210 710 9700
E7254X 0.6/
1.2
40/102 150 500/
1000
T,C 12 100 285 1180 9700
E7823FX 0.6/
1.2
31/85 150 500/
1000
T,C 14 100 285 1180 9700
E7864X 0.6/
1.2
36/96 150 400/
800
T,C 12 100 285 1200 9700
E7250X 0.6/
1.0
30/60 150 400/
700
T,C 16 100 285 1180 9700
E7869X 0.6/
1.2
40/100 150 500/
1000
T,C 12 100 420 1750 9700

♦Notes

  1. All tubes are neutral grounding; T -- Three-phase full-wave rectification; C -- Constant voltage
  2. Insert tubes

X-ray image intensifiers

X-ray image intensifiers
Type No. Useful
entrance
field
size
(mm)
Output
image
diameter
(mm)
Output
glass
thickness
(mm)
Central
Resolution
(Lp/cm)
Contrast ratio Conversion
factor
DQE
*3
(%)
10% 10mm *1 *2
E5881J-P1 95 20 7 77 22 18 90 10 42
E5877J-P1 95 20 7 77 22 18 90 10 42
50 110 - - - -
E5759SD-P1 140 15 5.5 48 23 17 350 41 65
E5863SD-P6A 140 20 7 54 30 20 160 18 65
E5883SD-P1A 140 20 7 54 30 20 180 21 65
105 64 - - - -
E5872J-P1 160 20 5.5 50 30 22 200 23 65
E5871J-P1 160 20 5.5 50 30 22 200 23 65
105 56 - - - -
E5804SD-P6 215 20 5.5 48 25 16 240 28 65
E5764SD-P3
E5764SD-P4
215 20 5.5 48 25 16 240 28 65
160 56 - - - -
120 66 - - - -
E5830SD-P4A 215 25 14 52 30 19 220 26 65
160 58 - - - -
120 68 - - - -
E5765SD-P2 290 25 5.5 46 24 15 250 29 65
215 50 - - - -
160 56 - - - -
E5796SD-P2 290 25 14 46 30 18 300 34 65
215 50 - - - -
160 56 - - - -
E5771J-P1 325 30 16 42 33 18 400 46 65
250 48 - - - -
170 56 - - - -
E5876SD-P2A 360 35 19 46 30 18 300 35 65
290 50 - - - -
215 56 - - - -
160 65 - - - -



♦Notes

  1. *1: cd/m2 /mR/s *2: cd/m2/µGy/s *3: IEC standard
  2. The information given herein may be modified without notice due to product improvement or further development.